logo

details van de producten

Created with Pixso. Thuis Created with Pixso. Producten Created with Pixso.
IN HET BIJZONDER Controlemechanisme
Created with Pixso.

Fluke 718 Series Pressure Calibrator | Industrial Pressure Calibration Instrument

Fluke 718 Series Pressure Calibrator | Industrial Pressure Calibration Instrument

Merknaam: Fluke
Modelnummer: 718 Reeksen
MOQ: 1
Prijs: Negotistical
Gedetailleerde informatie
Plaats van herkomst:
OEM
Meetbereik:
0 ma tot 24 ma
Oplossing:
0,001 doctorandus in de letteren
Nauwkeurigheid:
0,015% + 1 telling
Maximale spanning:
30 volt (niet in werking)
Temperatuur:
-40 °C tot 60 °C
Bedrijfstemperatuur:
-10 °C tot 55 °C
Verpakking Details:
Nieuwe originele doos
Productbeschrijving
Fluke 718 Series Pressure Calibrator

Industrial pressure calibration instrument designed for precision measurement and calibration applications.

Key Technical Specifications
Model Range Resolution Overpressure
Fluke 718 1G -1 psi to 1 psi
-68.9 mbar to 68.9 mbar
(-6.89 kPa to 6.89 kPa)
0.001 psi, 0.001 mbar 5X Full Scale
Fluke 718 30G -12 psi to 30 psi
-850 mbar to 2 bar
(-83 to 206.84 kPa)
0.001 psi, 0.1 mbar 2X Full Scale
Fluke 718 100G -12 psi to 100 psi
-850 mbar to 6.895 bar
(-83 to 689.48 kPa)
0.01 psi, 1 mbar 2X Full Scale
Fluke 718 300G -12 psi to 300 psi
-850 mbar to 20.68 bar
(-83 to 2068.42 kPa)
0.01 psi, 1 mbar 375 psi, 25 bar

Pressure Module Compatibility
Fluke 718 Series with 75OP Series Pressure Modules:
  • 48 pressure modules available
  • Range: 0 to 10000 psi (2.5 mbar to 700 bar with water medium)
  • Accuracy: 0.025% full scale (depending on module specifications)
  • Functions: 0, minimum, maximum, hold reading, damping
Additional Features
  • Built-in Air Pump: Functions include 0, minimum, maximum, hold reading, damping
  • Supported Pressure Units: psi, in H₂O (4°C), in H₂O (20°C), cm H₂O (4°C), cm H₂O (20°C), bar, mbar, kPa, inHg, mmHg, kg/cm²
  • Loop Power Supply: 24Vdc range with ±10% accuracy
  • Current Measurement: 0 to 24mA range with 0.001mA resolution and 0.015% +1 digit accuracy
Medium compatibility depends on pressure module specifications. Overpressure protection varies by model and configuration.